Monitoring system

Transducer excitation, amplifier, and filter

The sensing elements of most pressure transducers are four resistive elements connected as a Wheatstone bridge. Resistive bridges require the application of an excitation voltage (typically 3-5 V). Modern monitors provide very accurate (±0.1 per cent) excitation voltages to the transducers. These monitors also contain amplifiers which take the small voltages supplied by the transducer, typically 30 pV/mmHg, and 'magnify' them by about 1000 times. In amplifiers, 'magnification' is called 'gain'. With modern amplifiers, the gain is very stable and is calibrated with an accuracy to about ±1 per cent. Because of the accuracy, reliability, and stability of both the transducers and amplifiers, it is no longer necessary to calibrate pressure monitors ( Cooper..and...Pa.ulsen 1994; Gardner..1996). Most pressure amplifiers include a low-pass filter to filter out unwanted high-frequency signals.

Zero control

The zero control on a pressure monitor permits the adjustment of the output signal from the monitor to be zero when the fluid-air interface of a designated 'zeroing stopcock' is placed at the mid-axillary position. Once the zeroing process has been completed it is important that the vertical relationship between the patient and the pressure transducer remains fixed. If there is a vertical movement of the transducer or of the patient relative to the transducer, the system must be 're-zeroed.' However, the vertical position of the zeroing stopcock is unimportant once it is closed.

Sleep Apnea

Sleep Apnea

Have You Been Told Over And Over Again That You Snore A Lot, But You Choose To Ignore It? Have you been experiencing lack of sleep at night and find yourself waking up in the wee hours of the morning to find yourself gasping for air?

Get My Free Ebook


Post a comment